| Carbon Measurement | 0.00001% to 10.0000% |
| Sulfur Measurement | 0.00001% to 6.0000% |
| Relative Humidity | < 90% |
| Room Temperature | 10 to 30 °C |
| Sensitivity | 0.1 ppm |
| Reading Accuracy of Electronic Balance | 0.0001 g |
| Analysis Time | 35 seconds (can be set to 25 to 60 seconds) |
| High-Frequency Furnace | Power: 2.5 kVA; Frequency: 20 MHz |
| Oxygen Purity | ≥ 99.5% |
| Oxygen Input Pressure | 0.18 MPa ± 5% |
| Oxygen Load Pressure | 0.08 MPa ± 2% |
| Power Gas | Nitrogen or compressed air (purified) |
| Power Gas Input Pressure | 0.25 MPa ± 5% |
| Gas Flow | Top oxygen flow 1.0 to 2.0 L/min |
| Gas flow analysis | 3.0 to 4.5 L/min |
| Desiccant | Efficient color-changing desiccant |
| Filter agent | Quartz wool |
| Power supply | AC 220 V ± 5% |
Features
Applications
Infrared Carbon and Sulfur Analyzer is widely used for accurately determining carbon and sulfur content in various materials such as metals, minerals, and organic samples.
Still have questions?
If you cannot find answer to your question in our FAQ, you can always
contact us. We will answer shortly!
| Carbon Measuring Range | 0.0005% to 6.0000 % (w(C)) (extendable to 99.999%) |
| Sulfur Measuring Range | 0.0005% to 0.5000% (w(S)) (extendable to 99.999%) |
| Temperature Range | 10 to 30°C |
| Humidity Range | Less than 75% |
| Sample Weight | Standard 0.5g |