Alcohol Analyzer LB-10AA is capable of detecting breath alcohol levels ranging from 0 to 400 mg per 100 ml. It operates within an ambient temperature range of -5°C to 40°C and can be stored between -20°C and 70°C. It has a 32-bit processor, ensuring efficient and fast processing for precise test results. Our analyzer offers dual detection modes for both contact and non-contact testing for versatile use.
| Detection Range | 0 mg/100 ml to 400 mg/100 ml (breath alcohol concentration) |
| Resolution | 0.1 mg/100 ml |
| Indication Error | For C < 0.400: ± 0.06 mg/100 ml For 0.400 ≤ C < 1.000: ± 0.06 mg/100 ml For C ≥ 1.000: ± 6% |
| Indication Repeatability | For C < 0.400: ± 0.06 mg/100 ml For 0.400 ≤ C < 1.000: ± 0.06 mg/100 ml For C ≥ 1.000: ± 6% |
| Operating Temperature Range | - 5°C to 40°C |
| Storage Temperature Range | - 20°C to 70°C |
| Blowing Continuity Detection | 2.5 seconds (minimum flow rate: ≥ 20 L/min) |
| Alcohol Sensor Type | Electrochemical fuel cell type |
| Display | 128 × 160 color screen |
| Detection Modes | Contact and non-contact |
| Calibration Cycle | 6 months |
| Record Storage | Stores up to 100 test records |
| Charging Port | Android charging port |
| Battery | Rechargeable lithium battery |
| Dimensions (L × W × T) | 138 × 68 × 30 mm |
| Weight | 140 g |