Hematology Analyzer LB-10HEA can process up to 60 tests per hour, ensuring efficient and reliable blood cell analysis. It requires a maximum sample volume of 20 µL, ideal for limited or pediatric samples. It features a 10° incline design, enhancing visibility during manual observation. Our analyzer has 10.4-inch full-color LCD touchscreen, providing user-friendly navigation and clear display of test parameters.
Get Quote Download CatalogSpecifications
| Test Assay | 21 parameters + 3 histograms |
| Sample Volume | ≤ 20µL |
| Sampling Mode | Whole blood & pre-diluted blood |
| Throughput | 60 tests/hr |
| Test Parameters | WBC, LYM#, MID#, GRAN#, LYM%, MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC |
| Test Precision | WBC: ≤2.0% (3.5 to 15.0 × 10⁹/L) |
| Linear Range | WBC: 0 to 100.0 × 10⁹/L |
| Contamination Rate | WBC: ≤0.5% |
| Test Principle | Electric impedance method & Colorimetry |
| Operating Temperature range | 10°C to 35°C |
| Relative Humidity | 20% to 85% |
| Display | 10.4-inch LCD color touchscreen |
| Reagents | Built-in LH Lyse, External Diluent, Probe Cleaner |
| Quality Control | 3 Level QC, L-J graph, X-B |
| Calibration | Manual and Auto-calibration |
| Flag System | RBC, WBC, PLT abnormal samples, supports custom test groups |
| Communication Interface | 4 USB ports, 1 LAN port (HL7 protocol supported, compatible with LIS system) |
| Storage | Up to 50,000 results (histograms, scattergrams, patient info) |
| Power Consumption | 60W max |
| Power Supply | 100V to 240V AC, 50Hz / 60Hz |
| Product Dimension (L × W × H) | 390 × 285 × 390 mm |
| Weight | 16 kg |
Features
Low contamination rate
Automatic probe and tube cleaning
Automatic fault processing
Manual and automatic calibration
Built-in thermal printer
3-level quality control
Large data storage capacity
Low power consumption
Optional Accessories
20L Diluent
500 ml Lyse
500 ml Cleanser
Applications
Hematology Analyzer LB-10HEA is used for complete blood count testing, providing accurate measurement of blood cells and hemoglobin levels. It is widely used in hospitals, diagnostic laboratories, research centers, and pediatric clinics for routine and advanced hematology analysis.
| Processing Speed | 60 tests/hour |
| Sample Volumes | CBC + Diff mode: 20 µl, pre-diluted: 20 µl |
| Operating Temperature | 10 to 30 °C |
| Measuring Principle | Tri-angle laser scattering, flow cytometry for WBC differentiation, Impedance for RBC and PLT count, Cyanide-free method for HGB measurement |
| Processing Speed | CD: 80 tests/h CDR: 45 tests/h CD + ESR: 40 tests/h |
| Sample Volumes | CCD (Whole blood): 23 µl CD + ESR (Whole blood): 160 µl Predilute: 20 µl |
| Measuring Principles | WBC (IMG/Neu/Mon/Lym/Eos/Bas), NRBC/RET, PLT-H/PLT-O/IPF: SF Cube Cell Analysis Technology (Scatter, Fluorescence, 3D analysis) |
| Operating Temperature | 10℃ to 35℃ |
| Processing Speed | ≥ 55 tests/hour |
| Sample Volumes | Whole Blood Mode: 28 μL, Pre-diluted: 20 μL |
| Measuring Principle | Tri-angle laser scattering, Flow cytometry for WBC 5-part differentiation, Impedance for RBC & PLT count, Cyanide-free method for HGB measurement. |
| Precision Range | WBC: CV ≤ 2.0% (3.5 to 15.0) × 10⁹/L RBC: CV ≤ 1.5% (3.0 to 6.0) × 10¹²/L HGB: CV ≤ 1.5% (100 to 180 g/L) MCV: CV ≤ 1.0% (70 to 120 fL) PLT: CV ≤ 4.0% (150 to 500) × 10⁹/L |