Electrolyte Analyzer LB-10ELA supports whole blood and urine testing with a 160 µL sample volume, ensuring accurate and efficient electrolyte analysis. It utilizes the Ion Selective Electrode (ISE) method, providing accurate and ion-specific analysis. It features an integrated pilot lamp, ensuring clear indication of sample position and reagent liquid levels. Our analyzer has colourful touchscreen display, enabling precise monitoring and control.
$ 3600
Download Catalog Download ManualSpecifications
| Test Parameters | K⁺, Na⁺, Cl⁻, Ca²⁺, pH, TCO₂ |
| Calculated Parameters | nCa, TCa, AG |
| Measuring Method | Ion Selective Electrode Method |
| Blood Measuring Range | K⁺ - 0.50 to 15.00 mmol/L Na⁺ - 30.0 to 200.0 mmol/L Cl⁻ - 30.0 to 200.0 mmol/L Ca²⁺ - 0.10 to 5.00 mmol/L pH - 4.0 to 9.0 mmol/L TCO₂ - 4.0 to 70.0 mmol/L |
| Blood Accuracy | K⁺, Na⁺, Cl⁻: ≤ ±3.0% Ca²⁺: ≤ ±5.0% or 0.05 mmol/L pH: ≤ 0.15 TCO₂: ≤ 3.0 mmol/L |
| Blood Precision | K⁺, Na⁺, Cl⁻: ≤ 1.5% Ca²⁺: ≤ 1.5% pH: ≤ 1.0% TCO₂: ≤ 3.0% |
| Urine Measuring Range | K⁺ - 5.0 to 150.0 mmol/L Na⁺ - 30.0 to 200.0 mmol/L Cl⁻ - 10.0 to 300.0 mmol/L |
| Urine Accuracy | K⁺, Na⁺, Cl⁻: ≤ ±3.0% Ca²⁺: ≤ ±5.0% or 0.05 mmol/L pH: ≤ 0.15 TCO₂: ≤ 3.0 mmol/L |
| Urine Precision | K⁺, Na⁺, Cl⁻: ≤ 1.5% Ca²⁺: ≤ 1.5% pH: ≤ 1.0% TCO₂: ≤ 3.0% |
| Measuring Time | ≤100 seconds |
| Sample Capacity | 160 µL |
| Sample Type | Whole Blood and Urine |
| Sampling Method | Automatic |
| Operating Temperature Range | 15°C to 30°C |
| Humidity Level | ≤85% |
| Display | Colorful touchscreen display |
| Data Storage | Yes |
| Power Consumption | 50 W |
| Power Supply | AC 220V ± 22 V or AC 110V ± 11 V /, 50-60Hz |
| Product Dimension | 408 × 390 × 230 mm |
| Weight | 8 kg |
Features
Sample pump calibration
Automatic sampling
Air bubble detection
Automatic flushing
Solenoid valve system
User-friendly interface
Automatic calibration
Data storage capability
Standard Accessories
Bottle Cap × 3 pcs
Waste Reagent Cap × 1 pc
Waste Reagent Bottle × 1 pc
Pump Pipe × 2 pcs
Seal Ring of Reagent-Supply Port × 3 pcs
Printer Paper × 2 rolls
Nylon Fiber (500mm, φ0.6) × 1 pc
Spare Pipe (500mm, φ1.2) × 1 pc
Spare Pipe (500mm, φ1.5) × 1 pc
Spare Pipe (160mm, φ1.0) × 1 pc
2m Ground-Wire × 1 pc
Cable × 1 pc
Sampler System × 1 Set
Sample Cup × 50 pcs
Sample Turntable × 1 pc
Chassis of Sampler System × 1 pc
Pump Roller for Sample Pump × 5 pcs
Pump Roller Lube (1.5ml/piece) × 1 pc
Optional Accessories
Reagent | Volume |
Reagent A | 400 ml/bottle |
Reagent B | 200 ml/bottle |
Electrode Activation Reagent | 10 ml/bottle |
| Electrode Deproteinization Reagent | 10 ml/bottle |
R Reaction Reagent | 200 ml/bottle |
C Calibration Reagent | 15 ml/bottle |
Applications
Electrolyte Analyzer LB-10ELA is used for accurate measurement of key electrolytes in whole blood and urine. It is widely utilized in hospitals, clinical laboratories, emergency units, and biochemical research facilities.
Frequently asked questions For Electrolyte Analyzer
1. What is the primary function of a Labotronics Electrolyte Analyzer?
A Labotronics Electrolyte Analyzer is designed to measure key electrolytes such as sodium, potassium, chloride, and others in blood, serum, or plasma for accurate clinical diagnostics.
2. How do Labotronics Electrolyte Analyzers ensure accurate and ion-specific testing?
Labotronics Electrolyte Analyzers use advanced ion-selective electrode (ISE) and dry electrochemical technologies to deliver precise, ion-specific measurements across various biological samples.
3. Can Labotronics Electrolyte Analyzers handle multiple parameters in one test?
Yes, Labotronics electrolyte analyzers are capable of multi-parameter testing, allowing simultaneous measurement of several electrolytes in a single sample run.
4. How do Labotronics Electrolyte Analyzers help reduce contamination risks during testing?
Labotronics electrolyte analyzers feature closed-system operation with automated flushing and calibration, minimizing the chance of cross-contamination between samples.
5. Are Labotronics Electrolyte Analyzers portable for use in mobile or point-of-care settings?
Yes, select Labotronics electrolyte analyzers are compact and battery-powered, making them ideal for use in mobile clinics, fieldwork settings, and bedside diagnostics.
Still have questions?
If you cannot find answer to your question in our FAQ, you can always
contact us. We will answer shortly!
| Sample Types | Venous whole blood, plasma, and serum |
| Test Parameters | K+, Na+, Cl-, iCa2+, iMg2+ |
| Testing Principle | Electrochemistry/Microfluidic method |
| Blood Sample Capacity | 110 µL |
| Temperature Range | 10°C to 30°C |
| Sample Types | Venous whole blood, serum |
| Sample Volume Range | 80 µL to 120 µL |
| Throughput | Up to 55 samples in a row |
| Test Parameters | K⁺, Na⁺, Cl⁻, Ca²⁺, pH, Li⁺, tCO₂ |
| Testing Range | K+ - 1.00mmol/L to 16.0mmol/L Na+ - 30.0 mmol/l to 200.0 mmol/L Cl- - 30.0 mmol/l to 200.0 mmol/L Ca++ - 0.50 mmol/l to 5.00 mmol/L TCo2 - 5.0 mmol/l to 60.0 mmol/L pH - 6.00 to 9.00 Li + - 0.3 to 3mmol/L |
| Sample Type | Whole blood, serum, plasma, diluted urine |
| Sample Volume | 100 µL to 150 µL |
| Throughput | Up to 60 samples/hour |
| Measuring Range | K⁺: 0.50 to 15.00 mmol/L Na⁺: 30.0 to 200.0 mmol/L Cl⁻: 30.0 to 200.0 mmol/L Ca²⁺: 0.10 to 5.00 mmol/L pH: 4.0 to 9.50 (Unit) CO₂: 6.0 to 50.0 mmol/L |
| Precision | K⁺ ≤ 1.0% (CV%) Na⁺ ≤ 1.0% Cl⁻ ≤ 1.0% Ca²⁺ ≤ 3.0% pH ≤ 2.0% CO₂ ≤ 3.0% |
| Test Parameters | K, Na+, Cl-, Ca2+, pH(H+), Li+, TCO2 |
| Test Range | K+ - 1.00mmol/L to 16.0mmol/L Na+ - 30.0 mmol/l to 200.0 mmol/L Cl- - 30.0 mmol/l to 200.0 mmol/L Ca++ - 0.50 mmol/l to 5.00 mmol/L TCo2 - 5.0 mmol/l to 60.0 mmol/L pH - 6.00 to 9.00 Li + - 0.3 to 3mmol/L |
| Test Resolution | K+ - 0.01 mmol/L Na+ - 0.1 mmol/L Cl- - 0.1 mmol/L Ca++ - 0.01 mmol/L TCo2 - 0.1 mmol/L pH - 0.01 Li + - 0.01 mmol/L |
| Test Principle | Ion selective electrode |
| Test Method | Direct method |
Address: Labotronics Scientific. 1007 N Orange St.,
Wilmington, DE 19801, USA
Mail: info@labotronics.com