Octane Analyzer LB-10OAZ

Octane Analyzer LB-10OAZ

Octane analyzer LB-10OAZ is an analytical unit that automatically generates and print test results. Knock detection system ensures that the stability, repeatability and precision are less than 0.2 R octane. Microchip aids in rapid, accurate results. In compliance with ASTMD2699 and ASTMD2700 it evaluates gasoline type, and determines its Research Octane Number (RON) and Motor Octane Number (MON).

Buy Now Download Catalog

Specifications

Measurement range 40 RON ~ 120 RON
Speed range 900 ± 9r / min ( Motor method )
600 ± 6r / min ( Research method )
Pressure ≥ 0.15 MPa
Compression ratio range 4:1 ~ 18:1
Cylinder diameter 82.55 mm
Temperature of cylinder 100 ± 1 ℃
Humidity control range 3.56 ~ 7.12 grams water/kg
Motor method temperature ( adjustable ) 38 ℃ at ± 2 ℃
Research method temperature ( adjustable ) 52 ℃ at ± 1.0 ℃
Lubricating oil temperature 57 ℃ ± 8 ℃
Crankcase pressure -0.245 to -1.470 kPa
Mixture temperature ( Motor method ) 141 ℃ ~ 165 ℃ at ± 1 ℃
Cycle water temperature ≤ R.T + 10 ℃
Angle of ignition ( Motor method ) -24° ± 1.5°
Research method angle of ignition -13° ± 1.0°
Input valve size 0.20 mm
Exhaust valve size 0.25 mm
Spark plug gap 0.50 mm
Weight 900 kgs

Features

  • Measures leaded, unleaded gasoline and gasoline with different additives
  • Detection time: 15 minutes / gasoline sample
  • LCD touch screen to display real time data and results
  • Operation method: Interpolation and Compression ratio method
  • Equipped with SE air filtration system
  • Magnetic induction ignition for accurate tests
  • Cooling water circulation system
  • Low temperature test tank
  • FAD security protection system for safety of operators

Applications

Used in oil and petroleum industry, petrochemical industry, research institutes, laboratories, petroleum storage facilities, filling stations, customs houses, chemical industry, etc.

Our Team Of Experts Will Promptly Address Your Inquiry.

What our customers say

Fatima Jahaf logo

2 months ago

Manisha Mekhala logo

2 years ago

Morena Poli logo

2 years ago

Byrl H Hester logo

1 year ago

MIRZA AFZAL ALI logo

1 year ago

Beldine Ochieng logo

1 month ago

Rania Batta logo

1 year ago

Add a review

Nitrogen Analyzer is a scientific instrument designed to quantify the nitrogen content in various substances. Employing techniques such as combustion, chemiluminescence, or infrared spectroscopy, it accurately measures nitrogen levels in samples like soil, plant material, or organic compounds. Widely used in agriculture, environmental science, and industrial settings, the analyzer aids in optimizing fertilizer application, assessing soil health, and monitoring pollution. Its precision and speed make it a crucial tool for researchers and professionals, facilitating data-driven decisions and promoting efficient resource management.

Sample Types Solid, Gas, Liquid
Working Principle Chemiluminescence method (N)
Sample Injection Quantity Solid: 1to 20 mg Liquid: 5 to 20 μL Gas: 1to 5 mL
Measuring Range 0.05 to 20000 ppm (mg/L) (High concentration diluted)
Measuring range 79.0% to 99.9%
Sample Gas Flow 150 to 300 ml/min
Visible Display Window 90 × 25 mm
Accuracy ±2% F. S

A salt content analyzer is a specialized instrument used to precisely measure the concentration of salt or sodium chloride in various substances, particularly in food products and liquid solutions. These analyzers employ different methods, such as conductivity, titration, or ion-selective electrodes, to determine the amount of salt present. This data is crucial for quality control in the food industry, enabling manufacturers to ensure that products meet flavor, safety, and labeling standards. Salt content analyzers are also used in various other applications, including environmental monitoring and research, where accurate measurement of salinity in water samples is essential.

Halide concentration range 0.002 % ~ 0.02 %
Insulating resistance ≥ 2 MΩ
Extractor Potentiometric standard as per ASTM D6470
Heating power of extractor 300 W

Our Chlorine Content Analyzer is a specialized instrument used to quantitatively determine the chlorine content in various substances, particularly in chemical processes and industrial applications. It employs techniques such as titration, spectroscopy, or combustion analysis to measure the concentration of chlorine in samples. These analyzers are essential for quality control, process optimization, and compliance with environmental regulations in industries like petrochemicals, plastics, and water treatment. By providing accurate data on chlorine levels, these analyzers help ensure product quality, safety, and adherence to industry standards. Chlorine Content Analyzers play a crucial role in maintaining the integrity of processes and products across a range of industrial sectors, where precise chlorine content measurements are paramount.

Oxygen bomb pressure ≥ 20 MPa
Ignition voltage 12 V ~ 24 V
Repeatability 6 × 10-4 %
Ambient temperature 5 °C ~ 40 °C

A multi-test analyzer is a sophisticated and versatile laboratory instrument designed to perform a wide range of diagnostic tests and assays on clinical samples, such as blood, urine, or other bodily fluids. These instruments are equipped with multiple modules that can conduct various analyses, including chemistry, hematology, immunoassays, and coagulation studies. This comprehensive testing capability makes them invaluable in clinical laboratories and healthcare facilities, as they streamline the diagnostic process, enhance efficiency, and offer quick results for a range of medical conditions. The automation of sample handling, analysis, and result reporting minimizes human error, ensures precision, and accelerates patient diagnosis and treatment.

Function Chemistry + Coagulation
Sample Volume 50 µL to 500 µL
Wavelength Chemistry: 340 nm, 405 nm, 492 nm, 510 nm, 546 nm, 620 nm, 670 nm Coagulation: 450 nm, 670 nm
Incubator 2 reagent and 20 sample incubation positions, controlled at 37 °C ± 0.5 °C
Function Chemistry + Coagulation + Specific Protein
Sample Volume 50 µL to 500 µl
Wavelength Chemistry: 340 nm,405 nm , 492 nm, 510 nm, 546 nm, 620 nm, 670 nm Coagulation: 450 nm, 670 nm Specific protein: 450 nm, 670 nm
Incubator 2 reagent and 20 sample incubation positions , controlled at 37 °C ± 0.5 °C
Function Chemistry + Specific Protein
Sample Volume 100 µL to 500 µL
Wavelength Chemistry: 340 nm, 405 nm, 492 nm, 510 nm, 546 nm, 620 nm, 670 nm Specific Protein: 450 nm, 670 nm
Incubator 2 reagent and 20 sample incubation positions, controlled at 37 °C ± 0.5 °C

Our Portable Coal Content Analyzers offer fast and accurate analysis of coal properties in various field environments. These analyzers provide real-time measurement of key coal content parameters such as moisture, ash, and volatile matter. They are equipped with high-precision sensors to ensure reliable and consistent results. Their data storage capability allows for record-keeping and easy analysis of results. Their user-friendly interface simplifies operation, requiring minimal training. Our portable coal content analyzers feature a durable construction designed to withstand harsh field conditions, ensuring longevity and reliability.

Clean Coal Ash Content Error ≤ 1.0% (Low Ash Coal)
Clean Coal Calorimeter Error 120 cal/g
Clean Coal Moisture Error ≤ 1%
Thermal Coal Ash Content Error ≤ 2.5% (High Ash Coal)
WhatsApp
×

Get Quotation

Follow Us

Copyright © Labotronics Scientific 2025