Automatic Pensky-Martens Closed Cup Flash Point Tester LB-20PFP

Automatic Pensky-Martens Closed Cup Flash Point Tester LB-20PFP

Automatic Pensky-Martens Closed Cup Flash Point Tester LB-20PFP offers efficiency and precision. This cutting-edge instrument conforms to international benchmarks like ASTM D93 and ISO2719-88, ensuring reliable results. By employing an adaptive PID control algorithm, the tester dynamically adjusts the heating curve. With fully automated functionalities, it streamlines the testing process. Our commitment to innovation offers scientists an intuitive tool to enhance their analytical workflows.

Buy Now Download Catalog

Specifications

Ignition method Automatic
Measurement principle Thermocouple differential detection
Temperature Measurement Range Room temperature to 350 ℃
Resolution 0.1 ℃
Repeatability ±1 ℃ at flash point ≤ 110 ℃, ± 2 ℃ at flash point >110 ℃
Cooling Method Forced air cooling
Temperature sensor Imported platinum resistance (Pt100Ω)
Display Touch type color LCD display
Printer High-speed thermal printer, 56 mm paper width
Ambient conditions Temperature: 10 ℃ to 45 ℃, Humidity: 30% to 80%
Power Supply AC 220 V ± 10 %
Dimensions (L × W × H) 450 × 280 × 290 mm
Packing dimension (L × W × H) 530 × 400 × 450 mm
Net Weight 15 Kg
Gross weight 30 Kg

Features

  • Automatic Pensky-Martens Closed Cup Flash Point Tester with LCD display
  • PID control algorithm adjusts heating curve
  • Automatic stirring, ignition, detection, alarm, cooling, and printing functions
  • Safe airless source platinum alloy electric wire ignition method
  • Powerful data storage capacity for up to 300 records

Applications

Automatic Pensky-Martens Closed Cup Flash Point Tester is ideal for testing across petroleum, chemical, power, environmental protection, railway, and scientific research industries.

Request Quotation

Pensky-Martens Closed Cup Flash Point Tester

Ignition method Remote control
Ignition gas Pipe gas or LPG
Pressure ≤ 10 kPa
Measuring range 40 °C ~ 300 °C
Flame Pilot flame
Ignition method Automatic
Flame Pilot flame
Temperature sensor Glass pt100
Compliance ASTM D93
Ignition method Manual
Ambient Temperature 15 to 35 °C
Relative Humidity ≤ 85%
Compliance ASTM D93
Temperature Range Room Temperature to 350℃
Stirring Speed Grade A:125 r/min, Detection Temp Greater than 110℃
Temperature Control Accuracy ±0.1℃
Environmental Change Range 50℃ to 450℃