Standard gradient PCR machine LB-23GPCR is a Peltier based unit with sealed sample design. Mouse based operation discards traditional push button operation. The homogeneity of sample temperature is detected through 12 channelled temperature probes. The sealed sample design ensures a low temperature preservation and dry cleaning. Independent network connection discards complex operations and inconvenience of rational program updating. It is characterized with a remote-controlled software for diagnosis and data exchange of PCR machine for improved work efficiency.
Buy Now Download CatalogSpecifications
Number of wells | 384 |
Capacity per well | 0.2 mL |
Operational temperature range | 0 °C ~ 99.9 °C |
Gradient temperature range | 30 °C ~ 99 °C |
Gradient speed | 1 °C ~ 30 °C |
Gradient uniformity | ≤ 0.2 °C |
Heating and cooling rate range | 0.1 °C / s ~ 4.0 °C / s |
Maximum heating rate | ≥ 4 °C / s |
Maximum cooling rate | ≥ 4 °C / s |
Temperature uniformity | ≤ ± 0.2 °C |
Temperature accuracy | ≤ ± 0.1 °C |
Lid temperature | 20 °C ~ 110 °C |
Maximum number of cycles | 299 |
Ambient temperature | 5 °C ~ 30 °C |
Relative humidity | < 90 % |
Data storage program | 1000 |
Dimensions | 380 × 270 × 250 mm |
Weight | 7.8 kgs |
Features
Applications
Used for DNA amplification under gradient temperature system in the fields of genetics, immunology, genotyping, forensic, etc.Number of wells | 54 |
Capacity per well | 0.5 mL |
Operational temperature range | 0 °C ~ 99.9 °C |
Gradient temperature range | 30 °C ~ 99 °C |
Number of wells | 96 + 77 |
Capacity per well | 0.2 mL (96 wells) 0.5 mL (77 wells) |
Operational temperature range | 0 °C ~ 99.9 °C |
Gradient temperature range | 30 °C ~ 99 °C |
Number of wells | 96 |
Capacity per well | 0.2 mL |
Operational temperature range | 0 °C ~ 99.9 °C |
Gradient temperature range | 30 °C ~ 99 °C |
Address: Labotronics Scientific. 1007 N Orange St.,
Wilmington, DE 19801, USA
Mail: info@labotronics.com