Visible Spectrophotometer LB-41VSP is a light intensity measuring unit with wavelength range 320 nm ~ 1020 nm. Achromatic Czerny-Turner optical system and automatic light door function with manual wavelength setting ensures stable output. It is equipped with large sample chamber for accommodating cuvettes with tungsten lamp as light source increasing metering accuracy.
$ 1687.5
Download CatalogSpecifications
| Optical system | Achromatic Czerny-Turner (1200 L/mm grating) |
| Wavelength range | 350 nm ~ 1020 nm |
| Spectral bandwidth | 4 nm |
| Wavelength accuracy | ± 2 nm |
| Wavelength repeatability | 0.5 nm |
| Wavelength setting | Manual |
| Display | 4-bit LCD |
| Detector | Silicon photodiode |
| Stray light | 0.2 % T at 360 nm |
| Photometric accuracy | 0.3 % T |
| Photometric repeatability | 0.2 % T |
| Drift | 0.1 % T |
| Noise | ≤ 0.2 % T |
| Stability | ± 0.002 A/h at 500 nm |
| Light source | Tungsten |
| Measurement modes | T,A,C |
| Power | AC 110 V,50 Hz AC 220 V,60 Hz |
| Weight | 8 kgs |
Features
Applications
Used in biochemistry, clinical and research labs, petrochemical, agriculture, environmental protection for concentration measurement, sample analysis, impurities detection, chemical kinetics, etc.| Optical system | Double beam |
| Wavelength range | 320 nm ~ 1100 nm |
| Spectral bandwidth | 4 nm |
| Wavelength accuracy | ± 1 nm |
| Wavelength Range | 325 to 1000 nm |
| Spectral Bandwidth | 4 nm |
| Wavelength Accuracy | ±2.0 nm |
| Wavelength Repeatability | 1 nm |
| Optical system | Achromatic Czerny-Turner (1200 L/mm grating) |
| Wavelength range | 320 nm ~ 1100 nm |
| Spectral bandwidth | 2 nm |
| Wavelength accuracy | ± 0.5 nm |