| Temperature Parameters | Range: -5 to 110 ºC, 23 to 230 ºF Unit: ºC, ºF Resolution: 0.1 Accuracy: ±0.2 |
| Measurement Modes | Reading Mode: Auto Read (Fast, Medium, Slow), Timed, Continuous Reading Prompts: Reading, Stable, Locked, Temp. Compensation: ATC, MTC |
| Parameters | pH/ EC/ ISE/ Temp. (mV/ ORP/ pX/ Resistivity/ TDS /Sal.) |
| pH | Range: -2.00 to 20.00 ph Resolution: 0.1, 0.01 pH Accuracy: ±0.01 pH Calibration Points: Up to 5 Standard Recognition: NIST, GB, and DIN buffers Slope Limit |
| mV | Range: -2000.0 to 2000.0 mV Resolution: 0.1 mV Accuracy: ±0.3 mV or ±0.1% |
| pX | Range: -2.00 to 20.00 Resolution: 0.1, 0.01 pX Accuracy: ±0.01 pX Calibration Points: Up to 5 |
| ISE | Range: 1E-9 to 9.999E9 Unit: mol/L, mmol/L, g/L, mg/L, µg/L, ppm Resolution: Up to 4 significant digits Accuracy: ±0.5% Calibration Points: Up to 5 |
| Conductivity | Range: 0.000 µS/cm to 1000 mS/cm Resolution: 0.001 µS/cm minimum (varies with range selection), Accuracy: ±1.0% FS Reference Temperature: 20, 25 ºC Calibration Points: Up to 3 Standard Customization: 84 µS/cm, 1413 µS/cm, 12.88 mS/cm |
| Resistivity | Range: 5.00 Ω·cm to 20.00 MΩ·cm Resolution: 0.01 Ω·cm minimum Accuracy: ±1.0% FS |
| TDS | Range: 0.00 mg/L to 300 g/L Resolution: 0.01 mg/L Accuracy: ±1.0% FS |
| Salinity | Range: 0 to 8.00 % Resolution: 0.01% Accuracy: ±0.2% |
| Auto Shutdown Feature | Custom 1to 60 min, can be disabled |
| pH Electrode | BNC (Q9) |
| DO with Temperature Probe | 4-pin aviation connector |
| Conductivity with Temperature Probe | 5-pin aviation connector |
| Data Storage | 500 results each |
| Interface | USB 2.0 flash memory device, PC |
| Power Supply | AC Adapter, 100 to 240 V AC input, DC 9 V output |
| Dimension (L × W × H) | 242 × 195 × 68 mm |
| Weight | 900 g |
Features
Applications
Benchtop Multiparameter Meter is used for advanced pH and conductivity analysis to record pH, mV, ORP, pX, ISE, conductivity, total dissolved solids (TDS), salinity, resistivity, and temperature measurements across healthcare, environmental protection, biological agriculture and forestry, water quality analysis, scientific research and educational institutions.Still have questions?
If you cannot find answer to your question in our FAQ, you can always
contact us. We will answer shortly!
| COD Measuring Range | 10 to 15000 mg/L |
| COD Detection Method | Potassium Dichromate Method Potassium Permanganate method |
| COD Detection Lower Limit | 10 mg/L |
| COD Digestion Temperature & Time | 165 °C, 20 min |
| COD Accuracy | ≤±5% or ±4 mg/L |
| Measuring parameters | pH/EC/ISE/Temperature (mV/ORP/pX/Resistivity/TDS/Salinity) |
| Measuring range | pH: -2.000 to 20.000 pH mV/ORP: -2000.00 to 2000.00 mV pX range: -2.000 to 20.000 pX ISE: 1E-9 to 9.999E9 Conductivity: 0.000 µS/cm to 2000 mS/cm Resistivity: 5.00 Ω·cm to 100.00 MΩ·cm TDS: 0.000 to 1000 ppt Salinity: 0.0 to 80.0 ppt Temperature: -10 to 135°C, 14 to 275°F |
| Resolution | pH: 0.1, 0.01, 0.001 pH mV/ORP: 0.1, 0.01 mV pX: 0.1, 0.01, 0.001 pX ISE: Up to 4 significant units Conductivity: 0.001 µS/cm Resistivity: 0.01 Ω·cm TDS: 0.001 ppm Salinity: 0.1 ppt Temperature: 0.1°C/°F |
| Accuracy | pH: ±0.002 pH mV/ORP: ±0.1 mV or ±0.03% pX: ±0.002 pX ISE: ±0.3% Conductivity: ±0.5%FS Resistivity: ±0.5% FS TDS: ±0.5% Salinity: ±1 ppt Temperature: ±0.1 °C/°F |
| Calibration points | pH: Up to 6 ORP: 1 custom point (relative mV) pX: Up to 6 ISE: Up to 6 Conductivity: Up to 5 |