High Performance Liquid Chromatography(HPLC)
Our high-performance liquid chromatography (HPLC) is a powerful analytical technique used in chemistry and biochemistry to separate, identify, and quantify components within a liquid mixture. It operates by passing a liquid sample through a column packed with a stationary phase, which interacts with the sample's components based on their chemical properties. A pump delivers the sample into the column, while detectors, such as UV-Vis or fluorescence detectors, monitor the eluent as it exits the column. The resulting data generates chromatograms, providing information about the individual components' retention times, concentrations, and molecular characteristics. HPLC is valued for its high resolution, sensitivity, and versatility, making it an essential tool in pharmaceuticals, environmental analysis, food science, and biochemistry. It's used for tasks like drug purity testing, amino acid sequencing, and analyzing complex mixtures in fields ranging from quality control to scientific research, enabling precise and reliable analysis of diverse compounds and molecules in a wide array of applications.
Pump specifications |
Double plunger reciprocating pump |
Isocratic Pump (1 set) |
Flow range |
0.001 mL / min ̴ 10.000 mL / min |
Setting step size |
0.001 mL / min |
Flow accuracy |
≤ ± 0.3 % ( 1.00 ml / min, H2O , R.T. , 8.5 MPa ) |
Flow stability |
RSD ≤ 0.1 % ( 1.00 ml / min, H2O , R.T , 8.5 MPa ) |
Maximum pressure |
42 Mpa |
Display pressure error |
≤ ± 3.0 % |
Pressure fluctuation |
≤ 1.0 % |
Power dissipation |
80 W |
|
Detector specifications |
Wavelength repeatability |
± 0.1 nm |
Wavelength accuracy |
≤ ± 1.0 nm |
Spectrum band width |
8 nm |
Response time |
0.0 secs to 4.9 secs |
Linearity range |
1.8 AU ( 5 % ) |
Baseline noise |
≤ ± 1 × 10 - 5 AU |
Wavelength range |
190 nm ~ 700 nm |
Baseline drift |
≤ 2.0 × 10-4 AU / h |
Minimum detectable concentration |
4.0 × 10-9 g / mL ( Naphthalene in Methanol Solution ) |
Detector cell |
Optical Path - 10 mm |
|
Volume - 10 μl |
Light source |
Deuterium Lamp |
Power dissipation |
100 W |
Power supply |
AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
|
|
|
Pump specifications |
Double plunger reciprocating pump |
Binary high pressure gradient pump(2 set) |
Flow range |
0.001 mL / min ̴ 10.000 mL / min |
Setting step size |
0.001 mL / min |
Flow accuracy |
≤ ± 0.3 % ( 1.00 ml / min, H2O , R.T. , 8.5 MPa ) |
Flow stability |
RSD ≤ 0.1 % ( 1.00 ml / min, H2O , R.T , 8.5 MPa ) |
Maximum pressure |
42 Mpa |
Display pressure error |
≤ ± 3.0 % |
Pressure fluctuation |
≤ 1.0 % |
Power dissipation |
80 W |
|
Detector specifications |
Wavelength repeatability |
± 0.1nm |
Wavelength accuracy |
≤ ± 1.0 nm |
Spectrum band width |
8 nm |
Response time |
0.0 sec ̴ 4.9 sec |
Linearity range |
1.8 AU ( 5 % ) |
Baseline noise |
≤ ± 1×10 – 5 AU |
Wavelength range |
190 nm - 700 nm |
Baseline drift |
≤2.0 × 10 - 4 A U / h |
Minimum detectable concentration |
4.0×10-9 g / mL (Naphthalene in Methanol Solution) |
Detector cell |
Optical Path—10 mm |
|
Volume—10 μL |
Light source |
Deuterium Lamp |
Power dissipation |
100 W |
Power supply |
AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
|
|
|
Pump specifications |
Pump |
Isocratic Pump(1 Set) |
Flow rate range |
0.001 mL / min ̴ 10.000 mL / min |
Flow accuracy |
≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) |
Flow precision |
RSD ≤ 0.075 % (1.00 ml / min, H2O, R.T., 8.5 MPa) |
Maximum pressure |
45 MPa |
Pressure accuracy |
≤ ± 3.0 % |
Pressure pulsation |
≤ 0.1 MPa |
|
Detector specifications |
Wavelength range |
190 nm ~ 700 nm |
Wavelength repeatability |
≤ 0.1 nm |
Wavelength accuracy |
≤ ± 1.0 nm |
Band width |
8 nm |
Response time |
0.0 sec ̴ 5 sec |
Baseline noise |
≤ ± 0.5 × 10 – 5 AU |
Drift |
≤ 1.0 × 10 -4 AU / h |
Path length |
10 mm (analytical cell) |
Light source |
Deuterium Lamp |
|
Pump specifications |
Pump |
Binary high pressure gradient pump (2 Set) |
Flow Rate Range |
0.001 mL / min ̴ 10.000 mL / min |
Flow Accuracy |
≤ ± 0.2 % ( 1.00 ml / min, H2O, R.T., 8.5 Mpa ) |
Flow Precision |
RSD ≤ 0.075 % (1.00 ml / min, H2O, R.T., 8.5 Mpa ) |
Maximum Pressure |
45 MPa |
Pressure accuracy |
≤ ± 3.0 % |
Pressure pulsation |
≤ 0.1 MPa |
|
Detector specifications |
Wavelength range |
190 nm – 700 nm |
Wavelength repeatability |
≤ 0.1 nm |
Wavelength accuracy |
≤ ± 1.0 nm |
Band width |
8 nm |
Response time |
0.0 sec ̴ 5 sec |
Baseline noise |
≤ ± 0.5 ×10-5 AU |
Drift |
≤ 1.0 × 10-4 AU / h |
Path length |
10 mm (analytical cell) |
Light source |
Deuterium Lamp |
Power supply |
AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
|
Pump specifications |
Pump |
Quaternary low pressure gradient pump(1 Set) |
Flow rate range |
0.001 mL / min ̴ 10.000 mL / min |
Flow accuracy |
≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) |
Flow precision |
RSD ≤ 0.075 % (1.00 ml / min, H2O ,R.T., 8.5 MPa) |
Maximum pressure |
45 MPa |
Pressure accuracy |
≤ ± 3.0 % |
Pressure pulsation |
≤ 0.1 MPa |
|
Detector specifications |
Wavelength range |
190 nm ̴ 700 nm |
Wavelength repeatability |
≤ 0.1 nm |
Wavelength accuracy |
≤ ± 1.0 nm |
Band width |
8 nm |
Response time |
0.0 sec ̴ 5 sec |
Baseline noise |
≤ ± 0.5 × 10 - 5 AU |
Drift |
≤ 1.0 × 10-4 AU / h |
Path length |
10 mm (analytical cell) |
Light source |
Deuterium Lamp |
Power supply |
AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
|
|
|
Pump specifications |
Pump |
Quaternary low pressure gradient pump(1 Set) |
Flow rate range |
0.001 mL / min ̴ 10.000 mL / min |
Flow accuracy |
≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) |
Flow precision |
RSD ≤ 0.075 % (1.00 ml / min, H2O ,R.T., 8.5 MPa) |
Maximum pressure |
45 MPa |
Pressure accuracy |
≤ ± 3.0 % |
Pressure pulsation |
≤ 0.1 MPa |
|
Detector specifications |
Wavelength Range |
190 nm ̴ 800 nm |
Wavelength Repeatability |
≤ 0.1 nm |
Wavelength Accuracy |
≤± 1.0 nm |
Band Width |
6 nm |
Response Time |
0.0 s ̴ 5 s |
Baseline Noise |
≤ ± 0.25 × 10 -5AU (empty cell, 254 nm, 1.0 s) |
Drift |
≤0.5 × 10-4AU/h (empty cell, 254nm, 60min) |
Path length |
10 mm(analytical cell) |
Light Source |
Deuterium Lamp and Tungsten Lamp |
Power Supply |
AC110/220V±10%, 50/60Hz |
|
|
|
Pump specifications |
Type |
Gradient pump |
Flow Rate Range |
0.001 ̴ 10.000 ml / min, in 0.001 mL increments |
Flow Accuracy |
≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) |
Flow Precision |
RSD ≤ 0.06 % (1.00 ml / min, H2O, R.T., 8.5 MPa) |
Maximum Pressure |
63 MPa |
Pressure accuracy |
≤ ± 3.0 % |
Pressure pulsation |
≤ 0.1 MPa |
Composition rang |
0 % ̴ 100 % |
Composition precision |
≤ 0.2 % RSD |
|
Detector specifications |
Wavelength range |
190 nm ̴ 800 nm |
Wavelength repeatability |
≤ 0.1 nm |
Wavelength accuracy |
≤ ± 1.0 nm |
Band width |
6 nm |
Response time |
0.0 sec ̴ 5 sec |
Baseline noise |
≤ ± 0.25 × 10-5 AU (empty cell, 254 nm, 1.0 s) |
Drift |
≤0.5 × 10-4 A U / h (empty cell, 254 nm, 60 min) |
Path length |
10 mm (analytical cell) |
Light source |
Deuterium Lamp and Tungsten Lamp |
Power supply |
AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
|
|
|
Pump specifications |
Type |
Quaternary low pressure pump |
Flow rate range |
0.001 ̴ 10.000 mL / min, in 0.001 mL increments |
Flow accuracy |
≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) |
Flow precision |
RSD ≤ 0.06 % (1.00 ml / min, H2O, R.T., 8.5 MPa) |
Maximum pressure |
63 MPa |
Pressure accuracy |
≤ ± 3.0 % |
Pressure pulsation |
≤ 0.1 MPa |
Composition rang |
0 ̴ 100 % |
Composition precision |
≤ 0.3 % RSD |
|
Detector specifications |
Wavelength range |
190 nm ̴ 800 nm |
Wavelength repeatability |
≤ 0.1 nm |
Wavelength accuracy |
≤ ± 1.0 nm |
Band width |
6 nm |
Response time |
0.0 sec ̴ 5 sec |
Baseline noise |
≤ ± 0.25 × 10-5 AU (empty cell, 254 nm, 1.0 s) |
Drift |
≤ 0.5 ×10-4 AU / h (empty cell, 254 nm, 60 min) |
Path length |
10 mm (analytical cell) |
Light source |
Deuterium Lamp and Tungsten Lamp |
Power supply |
AC 110 V / 220 V ± 10%, 50 Hz / 60 Hz |
|
|
|